<Error>
<ErrorNumber>3002</ErrorNumber>
<ErrorFamily>Workstation Communications</ErrorFamily>
<ErrorMessage>
WSCOMM: Parameter syntax error
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>All</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The format of a workstation command is not recognized by the recorder. If the workstation is connected
and operating correctly, the recorder firmware and workstation software are not compatible.
</text></DescriptionText>
<DescriptionText><text>
The workstation should be checked to ensure that it is working properly.  Check that the serial interface
cable is connected to the correct port on the workstation.  If these are ok, the release notes for the
workstation software and recorder firmware should be checked for notes on compatiblity.  The firmware or
software should be upgraded accordingly.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Record error message and contact Creo Service
</text></DescriptionText>
</OperatorAction>
<Cause>
Workstation software malfunctioning
</Cause>
<Cause>
Serial interface cable connected to wrong port on workstation.
</Cause>
<Cause>
Workstation software and recorder firmware versions are not compatible.
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:32:18 2008 PST
</Updated>
<Links>
</Links>